If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 23, Number 2, 2018

High accuracy dosimetry with small pieces of Gafchromic films

Ryszard Dąbrowski, Izabela Drozdyk, Paweł Kukołowicz

Summary:

Aim

To investigate the influence of several factors on the accuracy of dose measurements and feasibility of application of small Gafchromic detectors for postal audit.

Background

Our experience showed that precision of dose measurements with small pieces of Gafchromic films may be significantly improved.

Materials and methods

Gafchromic films with dimensions of 1 × 1, 2 × 2 and 3 × 3 cm2 were exposed to 6 MV X-rays at dose levels of 50 cGy-210 cGy. The single- and multichannel methods (MM) were used for dose measurements. Detectors were scanned with an Epson V750PRO flatbed colour scanner. For 1 × 1 and larger detector sizes, separate calibration curves were established. The influence of the following factors was investigated: the heterogeneity of Gafchromic detectors group for single- and MM, ambient thermal detector conditions, the dose delivered on the measurement accuracy, application of two separate calibration curves for the smallest and larger pieces of films.

Results

The MM improves significantly the precision of dose measurement. The uncertainty attributed to detector active layer differences and scanner instabilities was about 1 cGy (1 StDev) regardless of dose and detector size. The ambient temperature of the environment in which films were stored after irradiation influenced the dose reading. Significant difference of transmission for detectors sized 1 × 1 and 2 × 2cm2 was observed. The maximal difference between applied dose and dose reading performed was 1.1%.

Conclusions

The MM with a scaling protocol leads to a very high precision of dose measurements. The ambient thermal detector environment causes significant changes of measured signal. The detector size has relevant impact on dose reading.

Signature: Rep Pract Oncol Radiother, 2018; 23(2) : 114-120