If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 22, Number 3, 2017

The semiconductor diode detector response as a function of field size and beam angle of high-energy photons

Tomasz Koper, Anna Kowalik, Sebastian Adamczyk

Summary:

Aim

The measurements of semiconductor diode detector response as a function of field size and beam angle of high-energy photons.

Background

In vivo dosimetry plays an important role in the therapeutic process of the patient. Because of the different orientation of the beam relative to the patient and different sizes of irradiation fields, it is extremely important to take into account the response of the detector depending on the angle and the size of the beam.

Materials and methods

In this study we used a 30 cm × 30 cm × 25 cm PMMA slab phantom. On the surface of the phantom, various semiconductor detectors were placed sequentially in two configurations, angle and tilt.

Results

For the measurements of the calibration factor based on the different value of the angle, the correction coefficient value was close to 1.00 for smaller values of the angle for all the detectors used in the energy range of 6–12 MV. For the measurements, the calibration factor based on the size of the field of irradiation to the value of the correction coefficient is 1.00 for the field of 8 cm × 8 cm and 10 cm × 10 cm. With the increase field size, the correction factor shows a linear relationship in the direction of value less than 1.00.

Conclusion

Flat Detectors – used for both photon beams generated by the accelerating potential of 6 MV and 20 MV show a greater angular dependence than the cylindrical detectors. Also, the repeatability of measurements made using the flat detector is less as evidenced by larger standard deviations for the results.

Signature: Rep Pract Oncol Radiother, 2017; 22(3) : 193-200