If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 6, Number 3, 2001

Verification of geometry reconstruction and dose calculation modules of the PLATO radiotherapy planning system

Góra E, Lesiak J, Barańczyk R, Rozwadowska-Bogusz B, Waligórski M.

Summary:

Purpose: Verification of reconstruction of application geometry and of calculations of dose distribution in the PLATO v. 13.7 system for brachytherapy treatment planning.
Materials and methods: The correctness of geometry reconstruction was verified using a Perspex phantom with lead markers. Dose calculations were verified by comparing values calculated by the system against values calculated "manually" and against published benchmark values.
Results: The average difference between marker coordinates in the phantom, as measured mechanically, and as reconstructed by the PLATO system, did not exceed 0.4 mm. The maximum relative difference between values of dose at given points, calculated by the PLATO system and values of dose calculated "manually" at those points, did not exceed 0.3%. The PLATO system was found to correctly reproduce the published benchmark data.
Conclusions: The geometry reconstruction software module unit and the dose distribution algorithm of the PLATO v. 13.7 planning system used at the Centre of Oncology in Kraków, were found to operate correctly.

Signature: Rep Pract Oncol Radiother, 2001; 6(3) : 141-147